LAUDA Integral IN 530 TW Process thermostat 400 V; 3/PE; 50 Hz & 460 V; 3/PE; 60

LAUDA Integral IN 530 TW Process thermostat 400 V; 3/PE; 50 Hz & 460 V; 3/PE; 60
The picture may be different from the article and can include supply, which is not in the standard scope.
 
Order-No.: L002667 Price on request for 1 piece plus VAT (19%) plus shipping costs
LAUDA Integral IN 530 TW Process thermostat
400 V; 3/PE; 50 Hz & 460 V; 3/PE; 60 Hz

Features:
Process thermostat with integrated cooling system for dynamic temperature control within external circuits
Coloured TFT display for simultaneous indication of actual & set values and graphic illustration of the temperature profile
Clear text menu navigation, six selectable languages DE, EN, FR, ES, IT, RU
Management of heat transfer liquids with stored properties
Easy input via cursor and soft keys. Additional Tmax key for overtemperature protection
SelfCheck Assistant for system diagnosis
Fully electronic continuous controller with PID action for internal & external control
Self adapt function for determination of control parameters
PowerAdapt system for the use of the maximum possible amount of heat permitted by the power supply system
Low-level and adjustable over-temperature protection with acoustic alarm for use with flammable and non-flammable liquids
Extremely powerful pressure pump
USB and Ethernet interface equipped as standard
Port for external Pt100 integrated, second external Pt100 feasible via interface module
Remote fault indication through floating contact
Option for upgrading up to 2 additional interfaces (RS 232/485, Profibus, analogue, contact or EtherCAT module)
Programmer with 150 temperature/time segments that can be separated into 5 programs
Adjustable bypass for pressure limiting
Digital display of pump pressure
Integrated web server for browser based operation in local area networks via PC, tablet or smart phone, secure data transfer due to authentication procedure and encryption
SmartCool system for energy-saving digital cooling management including compressor on-off control
Condenser cooling Water
Utilises traditional refrigerants (HFCs) in accordance with European legislation to control F-gases (EU) 517/2014

Technical Features (according to DIN 12876):
Working temperature range: -30 ... 120 °C
Ambient temperature range: 5 ... 40 °C
Temperature stability: 0.05 ± K
Heater power max.: 8 kW
Power consumption max.: 11 kW
Current max.: 16 A
Pump Pressure max.: 3,5 bar (50 Hz), 4,6 bar (60 Hz)
Pump flow rate max. (pressure): 40 L/min (50 Hz); 65 L/min (60 Hz)
In / Outlet connection thread (outside): M30 x 1,5
Inlet/outlet hose size: 3/43
Filling volume min.: 7.2 L
Filling volume max.: 20.5 L
Water cooling connection thread (outside): 3/4 3
Recommended cooling water temperature: 15 °C
Cooling water flow rate: 14.5 L/min
Recommended pressure difference cooling water: 3 bar
Pressure difference cooling water min.: 0.8 bar
Max. pressure difference cooling water: 10 bar
Maximal pressure cooling water: 10 bar
Overall dimensions (WxDxH): 560 x 550 x 1325 mm
Weight: 150 kg
Noise level: 62 dB(A)
Refrigerant stage 1: R-449A (GWP 1397); 1.200 kg; 1.7 t CO2-eq
Power supply: 400 V; 3/PE; 50 Hz & 460 V; 3/PE; 60 Hz
Power plug: Power cord with plug (IEC 60309, 5-pol, CEE, red, 16 A)

Cooling Capacity:
100 °C; Thermoöl; 6 kW (50 Hz); 6 kW (60 Hz)
20 °C; Ethanol; 6 kW (50 Hz); 6 kW (60 Hz)
10 °C; Ethanol; 5,5 kW (50 Hz); 5,5 kW (60 Hz)
0 °C; Ethanol; 4,5 kW (50 Hz); 4,5 kW (60 Hz)
-10 °C; Ethanol; 3 kW (50 Hz); 3 kW (60 Hz)
-20 °C; Ethanol; 1,6 kW (50 Hz); 1,6 kW (60 Hz)
-30 °C; Ethanol; 0,7 kW (50 Hz); 0,7 kW (60 Hz)

Standard Accessory:
2 nipples 3/4" with screw cap G3/4 for pump connectors
2 nipples 1/2" with screw cap G3/4 for cooling water
more